Solutions

Case9 For Better SpO2 Monitoring - Sensitivity Mode

The SpO2 value is often affected by the patient’s condition, artifacts created by external devices or by inappropriate attachment of probes, while monitoring continuously.
Nihon Kohden’s monitor has several features for better SpO2 measurement. Let us introduce one of the features here: Sensitivity Mode.
For other three features, please see in each page. Signal Quality Index, Pulse-amplitude Index, and Response.

Read More

Sensitivity Mode

Sensitivity mode allows you to improve the sensitivity of the pulse wave, therefore this feature is useful when it is difficult to detect the pulse of a patient with peripheral circulation insufficiency or when an IABP (Intra-Aortic Balloon Pumping) is used.
On the other hand, SpO2 value may be measured even if the probe is detached from the patient. Therefore, when selecting “Max” for the setting, it is necessary to confirm that the probe is appropriately attached to the patient.

Feature of setting the sensitivity of the pulse wave

Sensitivity mode on the display